Brief Introduction:
YS6002 benchtop color haze meter is developed by 3nh independently with proprietary intellectua property. This machine adopts transmittance optical geometry d/0 (diffused illumination, 0-degree viewing
angle) which confirms to CIE 15 standard. It can also realized the the optical geometry 0/0 (parallel illumination, 0-degree viewing angle).
YS6002 color haze meter uses 360~780nm composite full spectrum LED light source (which is closed to the spectrum of D65), and it can realize intervals 10nm spectrum by precision concave grating and 256 Image Element CMOS Sensor. Accordance with industry-standard, YS6002 will give the accuracy result for the measurement of haze, transmittance, chromaticity, color difference, WI, YI, Pt-Co Index and Gardner index etc. It is wildly used in the field of glass processing, plastic processing, film manufacturing, display screen processing and packaging industry, liquid pharmaceutical analysis, laboratory etc.
Features:
a. High configuration for the handware: 7 inch capacitive touch screen; 25000 data storage ; with USD / Bluetooth data port.
b. With high optical configuration and resolution, 154mm integrating sphere, concave grating, 256 Image Element CMOS Sensor, dual light path;
c. Adopt 360~780nm composite full spectrum LED light source (which is closed to the spectrum of D65), which is low power consumption and long life time.
d. Easy realize the precision measurement for 3 kinds of light sources A,C and D65 for haze (ASTM D1003/1044), transmittance, Pt-Co Index and Gardner index.
e. Multiple measurement modes : quality control mode and sample mode to meet more customers’ requirement;
f. Independent light source detector and built-in temperature sensor will continuously monitor the change of light sources and environment to ensure the measurement more reliable.
g. Observer angle 2/10, multiple light sources, multiple color spaces, multiple chromaticity to meet different kinds of customers’ testing requirements.
h. Good man-machine interactive to ensure the measurement accuracy and stable;
i. More powerful extended functions at the PC software.
Model | YS6002 |
Optical Geometry |
Transmittance:d/0 (diffused illumination, 0-degree viewing angle); similar to CIE15 standard 0/0(parallel illumination, 0-degree viewing angle); Confirms to the standard of ASTM D1003/1044, CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ASTM E308, DIN5033 Teil7 |
Application |
wildly used in the field of glass processing, plastic processing, film manufacturing, display screen processing and packaging industry, liquid pharmaceutical analysis, laboratory etc. |
Intergrating Sphere Size | Φ154mm |
Light Source | 360~780nm Combined LED Lamp |
Spectral Separation Device |
Concave-Grating |
Detector | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 360~780nm |
Wavelength Pitch | 10nm |
Half Bandwidth | 10nm |
Reflectance Range | 0~200% |
Measurement Aperture | Transmittance: Φ25.4mm |
Color Spaces | CIE LAB,XYZ,Yxy,LCh,CIE LUV, Musell,s-RGB, HunterLab,βxy,DIN Lab99 |
Color Difference Formulas |
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1), ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter) |
Other Colorimetric Data |
Haze (ASTM D1003/1044),transmittance rate T, WI(ASTM E313,CIE/ISO,AATCC,Hunter), YI(ASTM D1925,ASTM 313), TI(ASTM E313,CIE/ISO), Mt(Matamerism Index ), Color Strength, Staining fastness,Color Strength, Opacity, Pt-Co index and Gardner index |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,D55,D75, F1,F2,F3,F4,F5,F6,F7, F8,F9,F10,F11,F12 |
Display Data |
Spectral/Values, Chromaticity Values, Color Difference Values/Graph,Color Diagram, Color Simulation, Pass/ Fall Result, Color Offset |
Measurement Time | About 1.5s |
Repeatability |
Spectral transmittance: for test transmission reference white within ΔE*ab 0.05% (400~700nm: within 0.04%); Chromaticity Data: Φ25.4mm,within ΔE*ab 0.03 (after do warm-up and calibration, it is the average value to measure transmission reference white plate for 30 times at 5 second intervals) |
Inter-instrument Agreement |
Φ25.4mm,within ΔE*ab 0.15 (average value for 5 pcs standard transmittance color plates measurement ) |
Size | 370X300X200mm(L*W*H) |
Weight | About 9.6kg |
Power | DC 24V,3A Power Adapter |
Illuminant Life Span | 5 years, more than 3,000,000 times measurement |
Screen | 7-inch TFT Capacity Touch Screen Display |
Interface | USB,Printing Port |
Storage | Standard:5000; Sample:20000 |
Language | Simplified Chinese, Traditional Chinese, English |
Working Temperature | 0~40℃(32~104°F) |
Storage Temperature | -20~50℃(-4~122°F) |
Standard Accessory |
Power Adapter, User Manual, QC software(please download from website), USB cable, reflectance white and black board, transmittance black board, 25.4mm aperture, transmissive test component |
Optional Accessory | Miniature- printer |
Note | The specifications are subject to change without notice. |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Haze Meters
To answer this question, you first need to clarify your specific needs:
Is it for simple color quality control (QC)?
To coordinate with supply chain color management?
To monitor on-line color stability?
Do you require computer color matching or production color correction?
What is your budget?
Systematic supervision and evaluation of color performance in dyed products?
If you aren’t fully clear about your requirements at the initial stage, our color management engineers are ready to communicate with you. We will recommend solutions tailored to your needs. When conditions permit, we can provide prototype demonstrations and trial testing to ensure the instrument meets your practical application requirements.
Our machine is packed by standard export wooden box, it won’t be damaged. We’ve delivered many testing machines abroad by sea or by air without damage.
a. It’s needed to perform manual calibration for NH300 when first starting.
b. In the subsequent starting, no need to perform manual calibration.
c. Suggestion: it’s best to perform manual calibration once a week or when NH300 measured data is not accurate.
Threenh Technology colorimeters provide a color deviation prompt function according to the Lab value, which can provide a reference for color matching. The handheld colorimeter can only indicate blacker, whiter, redder, greener, yellower or bluer, but it cannot provide accurate color formulas. Only with the help of a spectrophotometer and professional color matching software can fully realized automated color matching.
Threenh Technology with excellent after sale service, Generally, Our Spectrophotometer instrument warranty is one years(Additional support period of 3 years). Within the warranty period, we can offer technique support and parts to parts change, necessarily, our engineers may go to your place for better support. Our instrument have professional After-sales Dept, if you have any technical problem or question with our goods welcome to contact us.